Principle
Potassium is a colorimeteric test.
The Indian Standard for drinking water does not specify an acceptable limit of potassium in water. However, according to the WHO, drinking water should have a potassium level below 12 mg/l.
Calibration
To prepare calibration standards, take 5 ml of distilled water. Depending on if you chose minimal or full calibration, prepare standards as below.
Full Calibration
Standard | Drops |
---|---|
0 mg/l | 0 |
10 mg/l | 10 |
20 mg/l | 20 |
30 mg/l | 30 |
Minimal Calibration
Standard | Drops |
---|---|
0 mg/l | 0 |
10 mg/l | 10 |
30 mg/l | 30 |
Take 5 ml of the 0 mg/l Potassium standard, dilute it to 10 ml with distilled water. Add 5 drops of Reagent A and mix well. Add 10 drops of Reagent B and mix well.
- Take the resulting coloured solution and fill it into the white cuvette provided.
- Place the cuvette under the light box and turn it on.
- Place the phone over the stand.
- Go to
Settings
>Colorimetric Test
>Calibrate
and choosePotassium
. - Select the calibration point, 0 mg/l in this case.
- Click on
Start Camera
. - Position the cuvette so that the middle, most uniformly lit portion is within the circle.
- Select
Analyze
. - You should have your first calibration point. Repeat similarly with the other calibration points.
Test
This test uses the Potassium reagents.
- Collect 5 ml of sample water, add 5 drops of Reagent A and 10 drops of Reagent B and shake for 5 seconds to mix well. Place the solution in the white cuvette.
- Place the cuvette under the lightbox and turn it on.
- Place the phone over the stand.
- From the home screen choose
Colorimetric Test
>Water
>Potassium
. - Choose
No Dilution
if this is the first test, you might have to dilute the sample and retry later if you get an error. - Click on
Start Camera
. - Position the cuvette so that the middle, most uniformly lit portion is within the circle.
- Select
Analyze
. - After some time, the app will display a result.