Principle
Chromium is a colorimeteric test. This test uses a 2, 5 diphenyl carbazide reagent.
The Indian Standard for drinking water holds that the acceptable limit is 0.05 mg/l and is not relaxable.
Calibration
To prepare calibration standards, take 5 ml of distilled water. Depending on if you chose minimal or full calibration, prepare standards as below.
Full Calibration
| Standard | Drops | 
|---|---|
| 0 mg/l | 0 | 
| 0.1 mg/l | 2 | 
| 0.5 mg/l | 10 | 
| 1 mg/l | 20 | 
| 2 mg/l | 30 | 
| 3 mg/l | 40 | 
Minimal Calibration
| Standard | Drops | 
|---|---|
| 0 mg/l | 0 | 
| 20 mg/l | 10 | 
| 40 mg/l | 20 | 
Take 5 ml of the 0 mg/l Chromium standard, and add 2 drops of Chromium Reagent A and 5 Drops of Chromium Reagent B and shake well.
- Take the resulting coloured solution and fill it into the white cuvette provided.
 - Place the cuvette under the light box and turn it on.
 - Place the phone over the stand.
 - Go to 
Settings>Colorimetric Test>Calibrateand chooseChromium. - Select the calibration point, 0 mg/l in this case.
 - Click on 
Start Camera. - Position the cuvette so that the middle, most uniformly lit portion is within the circle.
 - Select 
Analyze. - You should have your first calibration point. Repeat similarly with the other calibration points.
 
Test
This test uses the Chromium Reagent A and Chromium Reagent B.
- Collect 5 ml of sample water, add 2 drops of Chromium Reagent A, and 5 drops of Chromium Reagent B, and place in the white cuvette.
 - Place the cuvette under the light box and turn it on.
 - Place the phone over the stand.
 - From the home screen choose 
Colorimetric Test>Water>Chromium. - Choose 
No Dilutionif this is the first test, you might have to dilute the sample and retry later if you get an error. - Click on 
Start Camera. - Position the cuvette so that the middle, most uniformly lit portion is within the circle.
 - Select 
Analyze. - After some time, the app will display a result.